Detail of Recall

Type of Product:Device
Medsafe Reference:33819
Brand Name:Incubator, infant, stationary
Model:Giraffe OmniBed Carestation (2082844-001-XXX) [GTIN – 010084068211686221]
Affected:All Giraffe OmniBed Carestations manufactured between January 2018 and November 2021 and All Giraffe OmniBeds* and Giraffe OmniBed Carestations that have had heater doors replaced.
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose. If the doors become loose, the system activates a high priority alarm, and the canopy movement will stop.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/9/2024