Detail of Recall

Type of Product:Device
Medsafe Reference:33776
Brand Name:Angiographic X-Ray
Model:Philips Azurion, Allura Xper and MultiDiagnost Eleva
Affected:Multiple System Code
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:Azurion
Model:Philips Azurion, Allura Xper and MultiDiagnost Eleva
Affected:Multiple System Code
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified that due to the forces applied when moving or turning the monitor of the system, the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may become loose and/or break. If this issue occurs, plastic parts and/or metal bolts of the Cable Hose Carrier(s) may fall and/or part of the cable hose may drop following the detachment of a Cable Hose Carrier.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/9/2024