Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 33614 |
| Brand Name: | Anaesthesia Work Stations |
| Model: | Aisys CS2, Avance CS2, Avance CS2 Pro, Aisys, Avance, Amingo, Aespire 100, Aespire View, Aespire 7900, and Aespire 7100 anesthesia devices equipped with the Auxiliary Common Gas Outlet (ACGO) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Issue: | The manufacturer has become aware of a potential issue regarding the misconnection of the patient breathing circuit (circle) to the ACGO. The ACGO is used to provide fresh gas to an auxiliary manual breathing circuit (non-circle) that does not have CO2 absorbent capability. In the unlikely event that the inspiratory port of the circle patient breathing circuit is misconnected to the ACGO port, it can result in prolonged and excessive pressure leading to barotrauma. Note that the devices currently have a mechanical overpressure valve, user-adjustable airway pressure limits, and several alarms however if this situation is not identified and addressed by the attending clinician, the increase in breathing circuit pressure can be life threatening. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/8/2024 |