Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 33282 |
| Brand Name: | Ingenia, Achieva Series |
| Model: | Multiple Models |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Brand Name: | PMS Nederland MR Systems (MDR) |
| Model: | Multiple Models |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Issue: | The manufacturer has identified an issue where the g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room. The g-MDU, located in the technical room, is the single-entry point for the hospital electricity supply and distributes the electricity toward the various cabinets and components of the MR Scanner. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/6/2024 |