| Type of Product: | Device |
| Medsafe Reference: | 33232 |
| Brand Name: | MRI system, full-body, superconducting magnet |
| Model: | Multiple product names and software versions |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Healthcare Tianjin Company Ltd |
| Issue: | The manufacturer has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant. You may continue to use your MR system by following the manufacturer's instructions |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/6/2024 |