Detail of Recall

Type of Product:Device
Medsafe Reference:32508
Brand Name:Ingenia, Achieva Series
Model:
Affected:All MR systems with 60cm wide bore are affected
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified that the Quadrature Body Coil (QBC) seal adhesive may fail creating sharp edges that may come in contact with patients. The QBC seal may become loose as the patient table travels in a horizontal motion in and out of the system bore. The QBC seal is a rubber seal that is glued between the cone cover and QBC cover and functions to prevent sharp edges of the QBC cover from contacting patients during an examination.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/2/2024