Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:32239
Brand Name:Soltive Fibres
Model:TFL-FBX150S SOLTIVE LASER FIBRE 150 MICR- 200 MICR - 365 MICR, 550 MICR - 940 MICR, TFL-FBX150BS SOLTIVE LF 150M BALL TIP - 200M BALL TIP
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Gyrus ACMI Inc
Issue:The manufacturer conducted an internal review of the SOLTIVE Single-use Fibers and identified inconsistencies in the IFU. The IFU currently instructs the user that the SOLTIVE Single-use Laser Fiber is cleaved first and then stripped, leaving 20 mm of exposed fiber. This presents the potential for the fiber to break during use. The potential harms related to a broken fiber include burns, procedure delays, prolonged surgery, and foreign body in patient.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/12/2023