Type of Product:Device
Medsafe Reference:32175
Brand Name:Ingenia, Achieva Series
Model:Philips Achieva 1.5T, Achieva 1.5T Conversion, Ingenia 1.5T CX, Intera 1.5T, Intera 1.5T Power/Pulsar, SmartPath to dStream for 1.5T
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:16/08/2024: Update - As a reminder, the instructions within the original letter Section 4, Actions that should be taken by the
customer/user to prevent risks for patients or users are still advised and valid. Please continue to follow these
instructions until a correction is provided.

The manufacturer has identified an issue where a specific component failure in the Gradient Coil of the affected 1.5T MR Systems may act as a heat source with a potential to produce smoke and/or fire.
If the component failure occurs, the user may observe the following:
Repeated scan abort error messages due to gradient amplifier fault detection, described in Section 4.
Unusual noise from the system in the examination or technical room.
Smoke and/or fire from within the system.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/8/2024