Detail of Recall

Type of Product:Device
Medsafe Reference:31773
Brand Name:Portable ventilator, electric
Model:Astral 100/150
Affected:manufactured between 2013 and 2019
Software version:
Recalling Organisation:ResMed (NZ) Ltd
266 Runciman Road
Pukekohe
Auckland
Contact Information:0800 737 635
Manufacturer:ResMed Pty Ltd
Issue:The manufacturer is providing information on the second phase of corrective actions now available to affected Astral 100/150 ventilators, that have not already been corrected under phase one. Note, if a ventilator has had its circuit board replaced since the Product Correction announcement made in September 2023, it is no longer affected by this Product Correction. Phase Two: R6.2 Software Release. In the latest software release R6.2 (SX544-0603), Astral ventilators will now monitor the ongoing performance of the supercapacitor. If the software detects that the supercapacitor is degraded or malfunctioning, one of three new System Faults will activate on the ventilator with an audible alarm and message on the User Interface. These System Faults do not stop ventilation.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/3/2024