Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 31773 |
| Brand Name: | Portable ventilator, electric |
| Model: | Astral 100/150 |
| Affected: | manufactured between 2013 and 2019 |
| Software version: | |
| Recalling Organisation: | ResMed (NZ) Ltd 266 Runciman Road Pukekohe Auckland |
| Contact Information: | 0800 737 635 |
| Manufacturer: | ResMed Pty Ltd |
| Issue: | The manufacturer is providing information on the second phase of corrective actions now available to affected Astral 100/150 ventilators, that have not already been corrected under phase one. Note, if a ventilator has had its circuit board replaced since the Product Correction announcement made in September 2023, it is no longer affected by this Product Correction. Phase Two: R6.2 Software Release. In the latest software release R6.2 (SX544-0603), Astral ventilators will now monitor the ongoing performance of the supercapacitor. If the software detects that the supercapacitor is degraded or malfunctioning, one of three new System Faults will activate on the ventilator with an audible alarm and message on the User Interface. These System Faults do not stop ventilation. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/3/2024 |