Detail of Recall

Type of Product:Device
Medsafe Reference:31429
Brand Name:Angiographic X-Ray
Model:Model numbers - 459800772231, 459800772233, 459800772261, 459800772263, 459800415531, 459800415532, 459800415533 ,459800415534, 459800415535, 459800415571, 459800415572, 459800415573, 459800415574, 459800415575
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has become aware of several situations that can result in a loss of availability of the wireless foot switch during a procedure. Identified issues that may result in the wireless foot switch not being available are:
· Loss of Bluetooth connection due to interferences from other radio equipment
· Battery not fully charged
· Battery not holding its charge
· Damage in the charger, cable and/or connector of the wireless foot switch
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/7/2023