Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:31330
Brand Name:Ingenia, Achieva Series
Model:All MR System
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:PMS Nederland MR Systems (MDR)
Model:All MR Systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified an issue with certain MR room cage installations where the patient support (table) floor plate may have been incorrectly installed during construction of the MR suite. This may cause the table to become dislodged from the floor and potentially cause harm to the patient and/or operator. Note: The MR system will perform per its intended use. This issue is not related to a device malfunction. Update: The manufacturer is currently investigating a solution for identification of issue.
In the meantime, the manufacturer is providing this Updated which contains recommendations for continued use of the systems.


Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/4/2024