Detail of Recall

Type of Product:Device
Medsafe Reference:30901
Brand Name:Alaris System Infusion Pumps_CFN San Diego
Model:BD Alaris™ System PC Unit Model 8015. Alaris™ PC Unit with 802.11 b/g or 802.11 a/b/g wireless network card.
Affected:
Software version:Software version 12.1.X
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:CareFusion 303 Inc
Issue:The manufacturer is aware that when the customers are using an Alaris™ PC Unit with 802.11 b/g or 802.11 a/b/g wireless network card and software version 12.1.0 or 12.1.1 a failure to establish a network connection will occur for IP addresses that are in a specific range from x.x.x.224 to x.x.x.255. Alaris™ PC Unit 8015 with the 802.11 a/b/g/n wireless network card is not impacted by this issue.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/4/2023