Database updated: 3 May 2024

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29557
Brand Name:Anaesthesia Work Stations
Model:Aisys CS2 Anesthesia Systems: P/N: 1011-9050-000 – GTIN: 00840682102292 and Avance CS2 and Avance CS2 Pro Anesthesia Systems: P/N: 1009-9050-000 - GTIN: 00840682102322
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has identified that in certain Aisys CS2, Avance CS2, and Avance CS2 Pro Anesthesia Systems manufactured from January 1, 2021 to April 23, 2022 can potentially have the O2 & Air cylinder pressure transducers reversed. The anesthesia system can display the O2 cylinder pressure reading as the Air cylinder pressure and the Air cylinder pressure reading as the O2 cylinder pressure. The correct gases are connected to each supply line, but the screen will display the cylinder pressures as reversed.
In rare instances prolonged hypoxia could result if the oxygen cylinder became depleted when both air and oxygen pipeline gases are unavailable and both cylinders are being utilized but not measured appropriately.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/6/2022
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /