Detail of Recall

Type of Product:Device
Medsafe Reference:29091
Brand Name:Fore-Sight Elite Module Tissue Oximeter
Model:HEMFSM10, 01-06-3000
Affected:
Software version:
Recalling Organisation:Edwards Lifesciences (New Zealand) Limited
171 Morrin Road
St John
Auckland 1072
Contact Information:021 313 234
Manufacturer:Edwards Lifesciences LLC
Issue:The manufacturer has identified that the StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module (Model HEMFSM10) or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor (Model 01-06-3000) with the Fore-Sight Elite large sensor (FSESL) in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/3/2023