Detail of Recall

Type of Product:Device
Medsafe Reference:28617
Brand Name:MRI system, full-body, superconducting magnet
Model:Sagittal VIBRANT with NoSlabWrap factor greater than 1.0.
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Healthcare Tianjin Company Ltd
Issue:The manufacturer has recently become aware of an issue on the affected products listed below, which impacts breast imaging when running the Sagittal VIBRANT application with ASSET. The issue arises when the NoSlabWrap factor is set to a value larger than the default value of 1.0. In this scenario, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/11/2021