Database updated: 20 November 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28566 |
| Brand Name: | Anaesthesia Work Stations |
| Model: | Carestation 750/750c anesthesia delivery systems |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Issue: | The manufacturer has become aware of an issue where the Carestation 750/750c Anesthesia Delivery Systems - after a specific sequence of actions, audible tone of O2 supply pressure low alarm will remain silenced until system reboot. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/11/2021 |