Detail of Recall

Type of Product:Device
Medsafe Reference:28468
Brand Name:CRHF - CareLink SmartSync Device Manager System
Model:CareLink SmartSync™ Device Manager application software D00U005
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:The manufacturer has identified that the SmartSync interrogation sessions, or CareLink network transmissions may fail for Cobalt or Crome devices when the current session diagnostic data includes any VT/VF episode type with multiple therapy sequences and three or more data recording suspensions. For these specific episodes, the software is unable to decode and process the data. SmartSync will display a message indicating an Unexpected error occurred, and the application software requires restarting.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/10/2021