Detail of Recall

Type of Product:Device
Medsafe Reference:27546
Brand Name:Medtronic model 2090 programmer
Model:IPGs: Azure™ XT DR MRI SureScan™, Azure™ S DR MRI SureScan™, Azure™ XT SR MRI SureScan™, Azure™ S SR MRI SureScan™
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:CRDM - Carelink Encore Programmer Model 29901
Model:IPGs: Azure™ XT DR MRI SureScan™, Azure™ S DR MRI SureScan™, Azure™ XT SR MRI SureScan™, Azure™ S SR MRI SureScan™
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:CRDM - CareLink Network Device Data Management Application
Model:CRT-Ps: Percepta™ CRT-P MRI SureScan™, Percepta™ Quad CRT-P MRI SureScan™, Serena™ CRT-P MRI SureScan™, Serena™ Quad CRT-P MRI SureScan™
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:CRHF - CareLink SmartSync Device Manager System
Model:CRT-Ps: Percepta™ CRT-P MRI SureScan™, Percepta™ Quad CRT-P MRI SureScan™, Serena™ CRT-P MRI SureScan™, Serena™ Quad CRT-P MRI SureScan™
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue: The impedance measurement that is used should be based on the device's programmed polarity. However, for Orion and CRT-P Quad family of devices, the algorithm is coded to use the bipolar lead impedance measurement for the atrial and right ventricle leads, regardless of programmed polarity. This can result in the wrong impedance (and potentially an open-circuit impedance) being used to estimate longevity. This issue causes over-estimation of longevity in the first half of the battery's lifecycle.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:6/4/2021