Type of Product:Device
Medsafe Reference:27545
Brand Name:CRHF - CareLink SmartSync Device Manager System
Model:Percepta™ CRT-P MRI SureScan™, Percepta™ Quad CRT-P MRI SureScan and Serena™ CRT-P MRI SureScan™, Serena™ Quad CRT-P MRI SureScan
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:The manufacturer has identified a software issue with the Medtronic CareLink SmartSync Device Manager (SmartSync) that causes it to incorrectly inflate the battery longevity estimates for Percepta, Serena, and Solara Cardiac Resynchronization Therapy pacemaker (CRT-P) devices (referred to as the CRT-P Quad family of devices). This issue manifests in the period immediately preceding the device Recommended Replacement Time (RRT) when the device is in the pre-RRT state.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:6/4/2021