Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26085 |
| Brand Name: | Medtronic model 2090 programmer |
| Model: | 2090 CareLink™ Programmer |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Brand Name: | CRDM - Carelink Encore Programmer Model 29901 |
| Model: | 29901 Encore™ Programmer |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Brand Name: | CRDM - CareLink Network Device Data Management Application |
| Model: | CareLink Network Application Software 2491 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Brand Name: | CRHF - CareLink SmartSync Device Manager System |
| Model: | CareLink SmartSync™ Device Manager |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | There is a potential for Medtronic programmer and remote monitoring software applications to display an inaccurate remaining longevity estimate for a subset of implanted cardiac devices. It is important to note this is not a battery performance issue and there is no impact to device functionality. The Recommended Replacement Time (RRT) remains an accurate indicator for device replacement. Longevity estimates early after implantation and later in the device life are unaffected. The manufacturer is releasing the first of two waves of updates to correct this issue. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 3/6/2020 |