Detail of Recall

Type of Product:Device
Medsafe Reference:26007
Brand Name:Incubator, infant, stationary
Model:Certain Giraffe Incubators and Giraffe OmniBeds - Oxygen sensor compatibility with Giraffe Systems
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has become aware that certain Giraffe Incubators and Giraffe OmniBeds manufactured before 2012 can deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed in your device. The Servo Oxygen Module automatically regulates oxygen that is provided internally by the device. Standard pulse oximetry monitoring will detect patient oxygen level changes and therefore this issue should only result in temporary hypoxia or hyperoxia.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/5/2020