| Type of Product: | Device |
| Medsafe Reference: | 25797 |
| Brand Name: | Alaris System Infusion Pumps_CFN San Diego |
| Model: | |
| Affected: | All Product Codes |
| Software version: | |
| Recalling Organisation: | Becton Dickinson Ltd Unit 4 / 110 Mays Road Onehunga Auckland 1061 |
| Contact Information: | 0800 572 468 |
| Manufacturer: | CareFusion 303 Inc |
| Brand Name: | Alaris System Infusion pump, syringe_CFN San Diego |
| Model: | |
| Affected: | All Product Codes |
| Software version: | |
| Recalling Organisation: | Becton Dickinson Ltd Unit 4 / 110 Mays Road Onehunga Auckland 1061 |
| Contact Information: | 0800 572 468 |
| Manufacturer: | CareFusion 303 Inc |
| Issue: | System Error 255-XX-XXX can occur when a user selects two functions at the same time or in rapid succession (less than one second) or when not following typical workflows. This results in a synchronization issue between the PC unit and the modules. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/3/2020 |