Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25797
Brand Name:Alaris System Infusion Pumps_CFN San Diego
Model:
Affected:All Product Codes
Software version:
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:CareFusion 303 Inc
Brand Name:Alaris System Infusion pump, syringe_CFN San Diego
Model:
Affected:All Product Codes
Software version:
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:CareFusion 303 Inc
Issue:System Error 255-XX-XXX can occur when a user selects two functions at the same time or in rapid succession (less than one second) or when not following typical workflows. This results in a synchronization issue between the PC unit and the modules.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/3/2020