Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23121 |
| Brand Name: | MRI system, full-body, superconducting magnet |
| Model: | |
| Affected: | MRI systems Software Version issue |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | It has come to the manufacturers attention that the latest on-site software version may not have been reinstalled at some sites after service activities were performed that required software to be reloaded. If the incorrect version is loaded on the system, the images can be flipped left/right and/or there can be patient data mismatch. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/6/2018 |