Detail of Recall

Type of Product:Device
Medsafe Reference:18974
Brand Name:Intelerad InteleradPACS
Model:
Affected:Software version contains a "P" (eg PACS 4.5.1 P123) or contains an "R" with the following: PACS-4-5-1-R70, PACS-4-6-1-R80, PACS-4-8-1-R40, PACS-4-9-1-R30, PACS-4-10-1-R30, PACS-4-11-1-R20
Software version:Versions 4.5.1 to 4.11.1
Recalling Organisation:Pharmaceutical Marketing New Zealand Ltd
162 Nile Road
Milford
Auckland 0620
Contact Information:09 377 0660
Manufacturer:Intelerad Medical Systems Incorporated
Issue:Due to the asynchronous nature of image ingestion processing, a failure at the early stages of the ingestion process may occur leading to images not being visible to the end user, without any indication to the user that this has occurred.
The user may not be aware that images might be missing. A misdiagnosis or a delay in diagnosis and treatment can occur if health professionals base their diagnosis on incomplete image sets.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/8/2015