Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18563 |
| Brand Name: | Intelerad InteleradPACS |
| Model: | |
| Affected: | Software versions: 4.7.1 R27 and up, 4.8.1 R20 and up, 4.9.1 R17 and up, 4.10.1 R9 and up, 4.11.1 R3 and up |
| Software version: | |
| Recalling Organisation: | Pharmaceutical Marketing New Zealand Ltd 162 Nile Road Milford Auckland 0620 |
| Contact Information: | 09 377 0660 |
| Manufacturer: | Intelerad Medical Systems Incorporated |
| Issue: | Images successfully transmitted to IntelePACS may not be available for reading in all studies. This failure can be detected if the client has image count verification protocols. This failure can also be detected if complete series are missing as the user will know to look for those missing series. In addition if a large number of consecutive images are missing it will cause a noticeable gap in the anatomy, which is also highly detectable. However, under certain circumstances it can appear as if the study is complete with no user observable means to identify the failure. No patient incident has been reported to date. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/5/2015 |