Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14957 |
| Brand Name: | Medtronic MiniMed Insulin Infusion Pump Reservoirs |
| Model: | Models: MMT326A & MMT-332A |
| Affected: | Lots: H8442973, H8469703, H8496561, H8517079, H8463297, H8492449, H8510440, H8539013, H8467888, H8494645, H8512826 |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic MiniMed |
| Issue: | Affected lots of insulin reservoirs have an increased risk of leaking. A leak in the reservoir may result in delivery of less insulin than intended and, if there is an occlusion in the infusion set, the pump may not alarm. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Public Recall |
| Date Commenced: | 5/7/2013 |